Type (acronym) | rms |
---|---|
Type | reference measurement service |
Unique Nomination Number | C17RMS23 |
Reference Laboratory Status | Manufacturer
|
Analyte Name | magnesium |
IUPAC IFCC Number | CAS7439-95-4 |
Traceable to SI | Yes |
Principle of Analytical Procedure | Ion chromatography |
Reference Method Listed in JCTLM | Yes |
JCTLM Database Identifier Number | NRMeth 17 |
Procedure Difference in Your Laboratory | No |
Quantity | Amount-of-substance concentration |
From (0) | 0.507 |
To (0) | 2.014 |
Unit (0) | mmol/L |
From (1) | 2.2 |
To (1) | 2.2 |
Unit (1) | % |
Level of Confidence (%) | 95 % |
Comments on Concentration Range | The measurement range of concentration does not involve the dilution of the sample. |
Comments on Uncertainty Range | The expanded uncertainty is a relative uncertainty |
Survey Organisation | JCTLM(IFCC),www.dgkl-rfb.de:81 |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | www.dgkl-rfb.de:81/index.shtml |
Reference to Published Comparison Results | RELA-IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine. |
Additional Information 1 | ±3.75% |
Organization Acronym | AutoRF |
Organization Name | Reference laboratory of Beijing, Autobio Biochemistry Co., Ltd |
Organization Website | http://www.autobio-bj.com.cn/ |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Electrolytes |