Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C17RMS23 - magnesium in Blood serum

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC17RMS23
Reference Laboratory Status
Manufacturer
Analyte Namemagnesium
IUPAC IFCC NumberCAS7439-95-4
Traceable to SIYes
Principle of Analytical ProcedureIon chromatography
Reference Method Listed in JCTLMYes
JCTLM Database Identifier NumberNRMeth 17
Procedure Difference in Your LaboratoryNo
QuantityAmount-of-substance concentration
From (0)0.507
To (0)2.014
Unit (0)mmol/L
From (1)2.2
To (1)2.2
Unit (1)%
Level of Confidence (%)95 %
Comments on Concentration RangeThe measurement range of concentration does not involve the dilution of the sample.
Comments on Uncertainty RangeThe expanded uncertainty is a relative uncertainty
Survey OrganisationJCTLM(IFCC),www.dgkl-rfb.de:81
Intercomparison Results Publicly AvailableYes
Link to Comparison Resultswww.dgkl-rfb.de:81/index.shtml
Reference to Published Comparison ResultsRELA-IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine.
Additional Information 1±3.75%
Organization AcronymAutoRF
Organization NameReference laboratory of Beijing, Autobio Biochemistry Co., Ltd
Organization Websitehttp://www.autobio-bj.com.cn/
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameElectrolytes