JCTLM Database: higher-order reference materials, methods and services
Details : C17RMS45 - glucose in Blood plasma, Blood serum, Calibration solution
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C17RMS45 |
| Reference Laboratory Status | Manufacturer |
| Analyte Name | glucose |
| IUPAC IFCC Number | CAS50-99-7 |
| Applicable Matrices | Calibration solution, blood serum, blood plasma |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | Spectrophotometry |
| Reference Method Listed in JCTLM | Yes |
| JCTLM Database Identifier Number | NRMeth 34 |
| Procedure Difference in Your Laboratory | Yes |
| Quantity | Amount-of-substance concentration |
| From (0) | 2 |
| To (0) | 36.7 |
| Unit (0) | mmol/L |
| From (1) | 2.8 |
| To (1) | 1.2 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Comments on Uncertainty Range | The expanded uncertainty is a relative uncertainty. |
| Survey Organisation | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine |
| Intercomparison Results Publicly Available | Yes |
| Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
| Reference to Published Comparison Results | The code of this laboratory in RELA-IFCC EQAS is 63. |
| Additional Information 1 | within limits of equivalence |
| Validation Report for Improved Method | C17 RMS45 Validation report.pdf |
| Organization Acronym | LEADMAN |
| Organization Name | Reference Laboratory of Beijing Leadman Biochemistry Co., Ltd |
| Organization Website | http://www.leadmanbio.com |
| Organization Country ISO Code | CN |
| Organization Country Name | China |
| Analyte Category Name | Metabolites and substrates |