JCTLM Database: higher-order reference materials, methods and services
Details : C17RMS45 - glucose in Blood plasma, Blood serum, Calibration solution
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C17RMS45 |
Reference Laboratory Status | Manufacturer |
Analyte Name | glucose |
IUPAC IFCC Number | CAS50-99-7 |
Applicable Matrices | Calibration solution, blood serum, blood plasma |
Traceable to SI | Yes |
Principle of Analytical Procedure | Spectrophotometry |
Reference Method Listed in JCTLM | Yes |
JCTLM Database Identifier Number | NRMeth 34 |
Procedure Difference in Your Laboratory | Yes |
Quantity | Amount-of-substance concentration |
From (0) | 2 |
To (0) | 36.7 |
Unit (0) | mmol/L |
From (1) | 2.8 |
To (1) | 1.2 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Uncertainty Range | The expanded uncertainty is a relative uncertainty. |
Survey Organisation | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Reference to Published Comparison Results | The code of this laboratory in RELA-IFCC EQAS is 63. |
Additional Information 1 | within limits of equivalence |
Validation Report for Improved Method | C17 RMS45 Validation report.pdf |
Organization Acronym | LEADMAN |
Organization Name | Reference Laboratory of Beijing Leadman Biochemistry Co., Ltd |
Organization Website | http://www.leadmanbio.com |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Metabolites and substrates |