Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C17RMS45 - glucose in Blood plasma, Blood serum, Calibration solution

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC17RMS45
Reference Laboratory StatusManufacturer
Analyte Nameglucose
IUPAC IFCC NumberCAS50-99-7
Applicable MatricesCalibration solution, blood serum, blood plasma
Traceable to SIYes
Principle of Analytical ProcedureSpectrophotometry
Reference Method Listed in JCTLMYes
JCTLM Database Identifier NumberNRMeth 34
Procedure Difference in Your LaboratoryYes
QuantityAmount-of-substance concentration
From (0)2
To (0)36.7
Unit (0)mmol/L
From (1)2.8
To (1)1.2
Unit (1)%
Level of Confidence (%)95%
Comments on Uncertainty RangeThe expanded uncertainty is a relative uncertainty.
Survey OrganisationRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Reference to Published Comparison ResultsThe code of this laboratory in RELA-IFCC EQAS is 63.
Additional Information 1within limits of equivalence
Validation Report for Improved MethodC17 RMS45 Validation report.pdf
Organization AcronymLEADMAN
Organization NameReference Laboratory of Beijing Leadman Biochemistry Co., Ltd
Organization Websitehttp://www.leadmanbio.com
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameMetabolites and substrates