Details : C18RM31 - Alanine aminotransferase (ALT) in buffer
Type (acronym) | rm2 |
Type | reference material |
Unique Nomination Number | C18RM31 |
Analyte Name | alanine aminotransferase (ALT) |
IUPAC IFCC Number | EC 2.6.1.2 |
List I, II or III | List I |
Basis for Traceability Claim | Procedurally-defined: IFCC primary reference measurement procedurefor the measurement of catalytic concentration of alanine aminotransferase, Schumann et al., Clin. Chem. Lab. Med., 2002, 40, 718-724
|
Applicable Matrices | recombinant form of human cytosolic ALT in a buffer (pH 7.5) containing among others bovine serum albumin and polysaccharides |
Identifier Number | ERM-AD454k/IFCC |
Name | Alanine aminotransferase (ALT) in buffer |
Certifying Organization | JRC |
Quantity | Catalytic activity concentration |
Level of Confidence (%) | 95 % |
Reference to Commutability Study Publication | Intended use statement: The main purpose of the material is to control the performance of the IFCC primary reference measurement procedure for the measurement of catalytic activity concentration of ALT at 37 °C. It is not recommended to use this material as calibrant for routine measurement procedures. (see certification report Chapter 9.5)
Commutability statement: If ERM-AD454k/IFCC would be used to calibrate routine measurement procedures, extended commutability studies should be performed with these procedures (See the certification report chapter 8.2 ). |
Recommended for Publication in the JCTLM List | YES |
Organization Acronym | EU - JRC |
Organization Name | European Commission - Joint Research Centre |
Organization Website | https://crm.jrc.ec.europa.eu/ |
Organization Country ISO Code | BE |
Organization Country Name | Belgium |
Analyte Category Name | Enzymes |