JCTLM Database: higher-order reference materials, methods and services
Details : C18RMS24 - creatine kinase (CK) in Blood plasma, Blood serum, Calibration solution
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C18RMS24 |
| Reference Laboratory Status | Manufacturer |
| Analyte Name | creatine kinase (CK) |
| IUPAC IFCC Number | EC 2.7.3.2 |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | Kinetic spectrophotometry |
| Reference Method Listed in JCTLM | Yes |
| JCTLM Database Identifier Number | NRMeth 65 |
| Procedure Difference in Your Laboratory | No |
| Quantity | Catalytic activity concentration |
| From (0) | 0.83 |
| To (0) | 20.65 |
| Unit (0) | μkat/l |
| From (1) | 2 |
| To (1) | 2 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Comments on Concentration Range | The measurement range of concentration does not involve the dilution of the sample.
The conversion factor for enzyme catalytic activity concentrations: 1 U/L = 0.01667 µkat/L. |
| Comments on Uncertainty Range | The expanded uncertainty is a relative uncertainty. |
| Survey Organisation | RELA - IFCC External
Quality assessment
scheme for Reference
Laboratories in
Laboratory Medicine |
| Intercomparison Results Publicly Available | Yes |
| Link to Comparison Results | www.dgkl-rfb.de:81/index.shtml
|
| Reference to Published Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine |
| Organization Acronym | Zybio Inc. |
| Organization Name | Reference System Department of Zybio Inc. |
| Organization Website | https://www.zybio.com/ |
| Organization Country ISO Code | CN |
| Organization Country Name | China |
| Analyte Category Name | Enzymes |