Details : C18RMS37 - 17α-hydroxyprogesterone in Blood plasma, Blood serum

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC18RMS37
Reference Laboratory Status
Manufacturer
Analyte Name17α-hydroxyprogesterone
IUPAC IFCC NumberCAS68-96-2
Applicable Matriceshuman serum/plasma; lyophilized, fresh or frozen
Traceable to SIYes
Principle of Analytical ProcedureID-LC-MS/MS
Reference Method Listed in JCTLMYes
JCTLM Database Identifier NumberNRMeth 39
Procedure Difference in Your LaboratoryYes
QuantityAmount-of substance concentration
From (0)0.964
To (0)297
Unit (0)nmol/L
From (1)2.7
To (1)4.4
Unit (1)%
Level of Confidence (%)95%
Comments on Concentration RangeThe range of concentration does not involve the dilution of the sample
Survey OrganisationJCTLM(IFCC),www.dgkl-rfb.de:81
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/
Reference to Published Comparison ResultsThe code of this laboratory in RELA - IFCC EQAS is 134.
Additional Information 1The code of this laboratory in RELA - IFCC EQAS is 134
Validation Report for Improved MethodC18 RMS37 Validation report.pdf
Organization AcronymAutoRL Zhengzhou
Organization NameReference Laboratory of Autobio Diagnostics Co., Ltd
Organization Websitehttp://www.autobio.com.cn/
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameNon-peptide hormones