Details : C18RMS38 - 17β -estradiol in Blood plasma, Blood serum
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C18RMS38 |
Reference Laboratory Status | Manufacturer |
Analyte Name | 17β -estradiol |
IUPAC IFCC Number | CAS50-28-2 |
Applicable Matrices | human serum or plasma; fresh or frozen |
Traceable to SI | Yes |
Principle of Analytical Procedure | ID-LC/MS/MS |
Reference Method Listed in JCTLM | Yes |
JCTLM Database Identifier Number | C17RMP7 |
Procedure Difference in Your Laboratory | No |
Quantity | Amount-of-substance concentration |
From (0) | 0.117 |
To (0) | 13.49 |
Unit (0) | nmol/L |
From (1) | 3.7 |
To (1) | 2.2 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Uncertainty Range | The expanded uncertainty is relative uncertainty |
Survey Organisation | RELA - IFCC External
Quality assessment
scheme for Reference
Laboratories in
Laboratory Medicine |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | www.dgkl-rfb.de:81/index.shtml
|
Reference to Published Comparison Results | RELA-IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine.
The code of this laboratory in RELA - IFCC EQAS is 51. |
Organization Acronym | Maccura |
Organization Name | Maccura Biotechnology Co., Ltd. Reference System Dept. |
Organization Website | https://www.maccura.com/en/ |
Organization Country ISO Code | CN |
Organization Country Name | China |