JCTLM Database: higher-order reference materials, methods and services
Details : C18RMS4 - gamma-glutamyltransferase (GGT) in Blood serum
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C18RMS4 |
Reference Laboratory Status | Manufacturer |
Analyte Name | gamma-glutamyltransferase (GGT) |
IUPAC IFCC Number | EC 2.3.2.2 |
Applicable Matrices | Human serum or plasma; lyophilized, fresh or frozen |
Traceable to SI | Yes |
Principle of Analytical Procedure | Kinetic spectrophotometry |
Reference Method Listed in JCTLM | Yes |
JCTLM Database Identifier Number | NRMeth 69 |
Procedure Difference in Your Laboratory | No |
Quantity | Catalytic activity concentration |
From (0) | 0.322 |
To (0) | 4.14 |
Unit (0) | μkat/L |
From (1) | 2.2 |
To (1) | 2.1 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Concentration Range | The measurement range of concentration does not refer to the dilution of the sample. |
Comments on Uncertainty Range | The uncertainty components induced by imprecision, light path, wavelength, absorbance, pH, temperature, volumn,reagent batch number, reagent purity, method linearity and reagent aging are considered. |
Survey Organisation | IFCC Ring Trials for Reference Laboratories are organized on behalf of the IFCC by
the Reference Institute of Bioanalysis (RfB) of the German Society of Clinical Chemistry
and Laboratory Medicine (DGKL). |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | www.dgkl-rfb.de:81/index.shtml
|
Reference to Published Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine |
Organization Acronym | ReNJ |
Organization Name | Reference Laboratory of Nantong Jinglan Biotech Co., Ltd. |
Organization Website | http://www.ntjlbio.com/ |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Enzymes |