Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C18RMS5 - alpha-amylase (AMY) in Blood serum

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC18RMS5
Reference Laboratory StatusManufacturer
Analyte Namealpha-amylase (AMY)
IUPAC IFCC NumberEC 3.2.1.1
Applicable MatricesHuman serum or plasma; lyophilized, fresh or frozen
Traceable to SIYes
Principle of Analytical ProcedureKinetic spectrophotometry
Reference Method Listed in JCTLMYes
JCTLM Database Identifier NumberNRMeth 70
Procedure Difference in Your LaboratoryNo
QuantityCatalytic activity concentration
From (0)0.224
To (0)11.73
Unit (0)μkat/L
From (1)2.5
To (1)2.1
Unit (1)%
Level of Confidence (%)95%
Comments on Concentration RangeThe measurement range of concentration does not refer to the dilution of the sample.
Comments on Uncertainty RangeThe uncertainty components induced by imprecision, light path, wavelength, absorbance, pH, temperature, volumn,reagent batch number, reagent purity, method linearity and reagent aging are considered.
Survey OrganisationIFCC Ring Trials for Reference Laboratories are organized on behalf of the IFCC by the Reference Institute of Bioanalysis (RfB) of the German Society of Clinical Chemistry and Laboratory Medicine (DGKL).
Intercomparison Results Publicly AvailableYes
Link to Comparison Resultswww.dgkl-rfb.de:81/index.shtml
Reference to Published Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine
Organization AcronymReNJ
Organization NameReference Laboratory of Nantong Jinglan Biotech Co., Ltd.
Organization Websitehttp://www.ntjlbio.com/
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameEnzymes