JCTLM Database: higher-order reference materials, methods and services
Details : C18RMS54 - creatinine in Blood plasma, Blood serum
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C18RMS54 |
Reference Laboratory Status | Manufacturer |
Analyte Name | creatinine |
IUPAC IFCC Number | CAS60-27-5 |
Applicable Matrices | Blood serum or plasma |
Traceable to SI | Yes |
Principle of Analytical Procedure | ID-LC/MS |
Reference Method Listed in JCTLM | Yes |
JCTLM Database Identifier Number | C7RMP26 |
Procedure Difference in Your Laboratory | YES |
Quantity | Amount-of-substance concentration |
From (0) | 30 |
To (0) | 2093 |
Unit (0) | μmol/L |
From (1) | 1.8 |
To (1) | 1.8 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Concentration Range | The range of concentration does not involve the dilution of the sample |
Comments on Uncertainty Range | The expanded uncertainty is a relative uncertainty |
Survey Organisation | IFCC Ring Trials for Reference Laboratories are organized on behalf of the IFCC by the Reference Institute of Bioanalysis (RfB) of the German Society of Clinical Chemistry and Laboratory Medicine (DGKL), they are announced on the DGKL-RfB – website (www.dgkl-rfb.de:81/index.shtml) |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | www.dgkl-rfb.de:81/index.shtml
|
Reference to Published Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine |
Validation Report for Improved Method | Method validation report for the measurement of creatinine C18RMS54 2023.pdf |
Organization Acronym | KHB |
Organization Name | Reference Laboratory, Shanghai Kehua Bio-engineering Co.,Ltd |
Organization Website | https://www.skhb.com |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Metabolites and substrates |