Details : C18RMS8 - creatine kinase (CK) in Blood plasma, Blood serum, Calibration solution
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C18RMS8 |
Reference Laboratory Status | Manufacturer |
Analyte Name | creatine kinase (CK) |
IUPAC IFCC Number | EC 2.7.3.2 |
Traceable to SI | Yes |
Principle of Analytical Procedure | Kinetic spectrophotometry |
Reference Method Listed in JCTLM | Yes |
JCTLM Database Identifier Number | NRMeth 65 |
Procedure Difference in Your Laboratory | No |
Quantity | Catalytic activity concentration |
From (0) | 0.22 |
To (0) | 14.81 |
Unit (0) | µkat/l |
From (1) | 2.4 |
To (1) | 2.4 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Concentration Range | without dilution of the sample |
Reference Procedure Status Acceptable | / |
Survey Organisation | IFCC-RELA |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | www.dgkl-rfb.de:81/index.shtml
|
Reference to Published Comparison Results | RELA-IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine.
The code of this laboratory in RELA - IFCC EQAS is 195. |
Additional Information 1 | RELA labcode: 195 |
Remarks | SNIBE reference lab was established in 2021 and participated 2022RELA scheme for the first time.Besides, some enterprise calibrators were assigned value with the RMP. |
Organization Acronym | SNIBE |
Organization Name | Shenzhen New Industries Biomedical Engineering Co., Ltd. |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Enzymes |