Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C19RMS1 - glycated hemoglobin (HbA1c) in Whole blood

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC19RMS1
Reference Laboratory StatusHospital of University
Analyte Nameglycated hemoglobin (HbA1c)
IUPAC IFCC NumberQU60433
Traceable to SIYes
Principle of Analytical ProcedureLC-MS/MS
Reference Method Listed in JCTLMYes
JCTLM Database Identifier NumberC1RMP_P16_MS
Procedure Difference in Your LaboratoryNo
QuantityAmount-of-substance fraction [HbA1c/(HbA1c+HbA0)]
From (0)28.6
To (0)140
Unit (0)mmol/mol
From (1)1.4
To (1)1.4
Unit (1)%
Level of Confidence (%)95%
Comments on Concentration RangeThe measurement range spans from without dilution of the sample.
Comments on Uncertainty RangeThe expanded uncertainty depends on the matrix of the sample.
Survey OrganisationIFCC Ring Trials for Reference Laboratories are organized on behalf of the IFCC by the Reference Institute of Bioanalysis (RfB) of the German Society of Clinical Chemistry and Laboratory Medicine (DGKL).
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/
Reference to Published Comparison ResultsThe code of this laboratory in RELA - IFCC EQAS is 65
Additional Information 1No. 65 is the number of our lab participation in RELA - IFCC External Quality assessment scheme for Reference Laboratories
RemarksIncluding PT measurement provided by China National Center for Clinical Laboratories and RELA - IFCC External Quality assessment scheme for Reference Laboratories, and lab assigned value for Guangdong Center for Clinical Laboratories.
Organization AcronymGPHCM
Organization NameGuangdong Provincial Hospital of Chinese Medicine
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameProteins