JCTLM Database: higher-order reference materials, methods and services
Details : C19RMS2 - total protein in Blood serum
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C19RMS2 |
| Reference Laboratory Status | Manufacturer |
| Analyte Name | total protein |
| IUPAC IFCC Number | total protein |
| Applicable Matrices | serum or serum like material; fresh, frozen or lyophilized |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | Biuret reaction |
| Reference Method Listed in JCTLM | Yes |
| JCTLM Database Identifier Number | C17RMP3 |
| Procedure Difference in Your Laboratory | No |
| Quantity | Mass concentration |
| From (0) | 30.6 |
| To (0) | 130 |
| Unit (0) | g/L |
| From (1) | 2.5 |
| To (1) | 2.5 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Comments on Uncertainty Range | The expanded uncertainty is relative uncertainty. |
| Survey Organisation | RELA - IFCC External
Quality assessment
scheme for Reference
Laboratories in
Laboratory Medicine |
| Intercomparison Results Publicly Available | Yes |
| Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/ |
| Reference to Published Comparison Results | The code of this laboratory in RELA - IFCC EQAS is 51. |
| Organization Acronym | Maccura |
| Organization Name | Maccura Biotechnology Co., Ltd. Reference System Dept. |
| Organization Website | https://www.maccura.com/en/ |
| Organization Country ISO Code | CN |
| Organization Country Name | China |
| Analyte Category Name | Proteins |