JCTLM Database: higher-order reference materials, methods and services
Details : C19RMS28 - lactate dehydrogenase (LDH) in Blood plasma, Blood serum
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C19RMS28 |
| Reference Laboratory Status | Manufacturer |
| Analyte Name | lactate dehydrogenase (LDH) |
| IUPAC IFCC Number | EC1.1.1.27 |
| Applicable Matrices | Human serum / plasma; lyophilized, fresh or frozen |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | Kinetic spectrophotometry |
| Reference Method Listed in JCTLM | Yes |
| JCTLM Database Identifier Number | NRMeth 66 |
| Procedure Difference in Your Laboratory | No |
| Quantity | Catalytic activity concentration |
| From (0) | 1.65 |
| To (0) | 9.57 |
| Unit (0) | μkat/L |
| From (1) | 2.7 |
| To (1) | 1.7 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Comments on Concentration Range | The concentration measurement range doesn't involve the dilution of the sample. |
| Comments on Uncertainty Range | The expanded uncertainty is a relative uncertainty. |
| Survey Organisation | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine |
| Intercomparison Results Publicly Available | Yes |
| Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/ |
| Additional Information 1 | The code of this laboratory in RELA-IFCC EQAS is 204. |
| Additional Information 2 | RELA 2022 and RELA 2023:within limits of equivalence |
| Organization Acronym | HIGHTOP Biotech |
| Organization Name | Qingdao HIGHTOP Biotech Co., Ltd. |
| Organization Website | http://www.hightopqd.com/ |
| Organization Country ISO Code | CN |
| Organization Country Name | China |
| Analyte Category Name | Enzymes |