Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C19RMS32 - alpha-amylase (AMY) in Blood plasma, Blood serum

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC19RMS32
Reference Laboratory StatusManufacturer
Analyte Namealpha-amylase (AMY)
IUPAC IFCC NumberEC3.2.1.1
Applicable MatricesHuman serum / plasma; lyophilized, fresh or frozen
Traceable to SIYes
Principle of Analytical ProcedureKinetic spectrophotometry
Reference Method Listed in JCTLMYes
JCTLM Database Identifier NumberNRMeth 70
Procedure Difference in Your LaboratoryNo
QuantityCatalytic activity concentration
From (0)1.11
To (0)11.04
Unit (0)μkat/L
From (1)3.8
To (1)3
Unit (1)%
Level of Confidence (%)95%
Comments on Concentration RangeThe concentration measurement range doesn't involve the dilution of the sample.
Comments on Uncertainty RangeThe expanded uncertainty is a relative uncertainty.
Survey OrganisationRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/
Additional Information 1The code of this laboratory in RELA-IFCC EQAS is 204.
Additional Information 2RELA 2022 and RELA 2023:within limits of equivalence
Organization AcronymHIGHTOP Biotech
Organization NameQingdao HIGHTOP Biotech Co., Ltd.
Organization Websitehttp://www.hightopqd.com/
Organization Country ISO CodeCN
Organization Country NameChina
Analyte Category NameEnzymes