Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C1RMS108 - aspartate aminotransferase (AST) in Blood serum, Blood plasma

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC1RMS108
Reference Laboratory StatusEQAS-Organiser
Analyte Nameaspartate aminotransferase (AST)
IUPAC IFCC NumberEC 2.6.1.1
Traceable to SIYes
Principle of Analytical ProcedureKinetic spectrophotometry
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureIFCC Ref. Meth 37°C - AST
JCTLM Database Identifier NumberNRMeth 68
Procedure Difference in Your LaboratoryYes
QuantityCatalytic activity concentration
From (0)0.33
To (0)6.67
Unit (0)µkat/l
From (1)2.5
To (1)2.5
Unit (1)%
Level of Confidence (%)95%
Comments on Concentration RangeThe conversion factor for enzyme catalytic activity concentrations: 1 U/L = 0.01667 µkat/L
Reference Procedure Status AcceptableSame uncertainty for all the measurements at every concentration: is it a real estimate? Trueness controls are not certified materials
Laboratory Preparing for AccreditationYes
Survey OrganisationJCTLM
Intercomparison Results Publicly AvailableYes - JCTLM-Website
Link to Comparison ResultsData from comparison studies are available on request
Participation Status in Intercomparison Surveys acceptableYES
RemarksMethod used for setting target values in national EQA-surveys since 1997
Recommended for Publication in the JCTLM ListYes
Organization AcronymINSTAND e. V.
Organization NameSociety for Promoting Quality Assurance in Medical Laboratories
Organization Websitehttp://www.instand-ev.de
Organization Country ISO CodeDE
Organization Country NameGermany
Analyte Category NameEnzymes