Details : C1RMS109 - γ-glutamyltransferase (GGT) in Blood serum, Blood plasma
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C1RMS109 |
Reference Laboratory Status | EQAS-Organiser |
Analyte Name | γ-glutamyltransferase (GGT) |
IUPAC IFCC Number | EC 2.3.2.2 |
Traceable to SI | Yes |
Principle of Analytical Procedure | Kinetic spectrophotometry |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | IFCC Ref. Meth 37°C - Gamma-GT |
JCTLM Database Identifier Number | NRMeth 69 |
Procedure Difference in Your Laboratory | Yes |
Quantity | Catalytic activity concentration |
From (0) | 0.33 |
To (0) | 5 |
Unit (0) | µkat/l |
From (1) | 2.5 |
To (1) | 2.5 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Concentration Range | The conversion factor for enzyme catalytic activity concentrations: 1 U/L = 0.01667 µkat/L |
Reference Procedure Status Acceptable | Same uncertainty for all the measurements at every concentration: is it a real estimate? |
Laboratory Preparing for Accreditation | Yes |
Survey Organisation | JCTLM |
Intercomparison Results Publicly Available | Yes - JCTLM-Website |
Link to Comparison Results | Data from comparison studies are available on request |
Participation Status in Intercomparison Surveys acceptable | YES |
Remarks | Method used for setting target values in national EQA-surveys since 1998 |
Recommended for Publication in the JCTLM List | Yes |
Organization Acronym | INSTAND e. V. |
Organization Name | Society for Promoting Quality Assurance in Medical Laboratories |
Organization Website | http://www.instand-ev.de |
Organization Country ISO Code | DE |
Organization Country Name | Germany |
Analyte Category Name | Enzymes |