JCTLM Database: higher-order reference materials, methods and services
Details : C1RMS118 - total protein in Blood serum, Blood plasma
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C1RMS118 |
| Reference Laboratory Status | EQAS-Organiser |
| Analyte Name | total protein |
| IUPAC IFCC Number | total protein |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | Biuret /Spectrophotometry |
| Reference Method Listed in JCTLM | ay |
| Name of JCTLM Listed Reference Method Procedure | IFCC DB entry number "NRMeth 30" |
| JCTLM Database Identifier Number | NRMeth 30 |
| Quantity | Mass concentration |
| From (0) | 25 |
| To (0) | 130 |
| Unit (0) | g/l |
| From (1) | 1.5 |
| To (1) | 1.5 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Reference Procedure Status Acceptable | Yes |
| Laboratory Preparing for Accreditation | Yes |
| Laboratory Status | Accreditation in Preparation |
| Survey Organisation | JCTLM |
| Intercomparison Results Publicly Available | Yes - JCTLM-Website |
| Link to Comparison Results | Data from comparison studies are available on request |
| Participation Status in Intercomparison Surveys acceptable | YES |
| Remarks | Method used for setting target values in national EQA-surveys since 2003 |
| Comments for JCTLM DB Publication | The primary calibrator used is a bovine albumin material. |
| Recommended for Publication in the JCTLM List | Yes |
| Organization Acronym | INSTAND e. V. |
| Organization Name | Society for Promoting Quality Assurance in Medical Laboratories |
| Organization Website | http://www.instand-ev.de |
| Organization Country ISO Code | DE |
| Organization Country Name | Germany |
| Analyte Category Name | Proteins |