JCTLM Database: higher-order reference materials, methods and services
Details : C1RMS118 - total protein in Blood serum, Blood plasma
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C1RMS118 |
Reference Laboratory Status | EQAS-Organiser |
Analyte Name | total protein |
IUPAC IFCC Number | total protein |
Traceable to SI | Yes |
Principle of Analytical Procedure | Biuret /Spectrophotometry |
Reference Method Listed in JCTLM | ay |
Name of JCTLM Listed Reference Method Procedure | IFCC DB entry number "NRMeth 30" |
JCTLM Database Identifier Number | NRMeth 30 |
Quantity | Mass concentration |
From (0) | 25 |
To (0) | 130 |
Unit (0) | g/l |
From (1) | 1.5 |
To (1) | 1.5 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Reference Procedure Status Acceptable | Yes |
Laboratory Preparing for Accreditation | Yes |
Laboratory Status | Accreditation in Preparation |
Survey Organisation | JCTLM |
Intercomparison Results Publicly Available | Yes - JCTLM-Website |
Link to Comparison Results | Data from comparison studies are available on request |
Participation Status in Intercomparison Surveys acceptable | YES |
Remarks | Method used for setting target values in national EQA-surveys since 2003 |
Comments for JCTLM DB Publication | The primary calibrator used is a bovine albumin material. |
Recommended for Publication in the JCTLM List | Yes |
Organization Acronym | INSTAND e. V. |
Organization Name | Society for Promoting Quality Assurance in Medical Laboratories |
Organization Website | http://www.instand-ev.de |
Organization Country ISO Code | DE |
Organization Country Name | Germany |
Analyte Category Name | Proteins |