Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C1RMS143 - glucose in Whole blood, Blood serum, Blood plasma, Other, Urine

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC1RMS143
Reference Laboratory StatusCalibration laboratory of the DGKL
Analyte Nameglucose
IUPAC IFCC NumberCAS50-99-7
Applicable Matriceswhole blood, blood serum, blood plasma, urine, cerebro-spinal fluid
Traceable to SIYes
Principle of Analytical ProcedureIsotope dilution gas chromatography mass spectrometry (ID/GC/MS)
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureNIST definitive method for serum glucose
JCTLM Database Identifier NumberNRMeth 80
Procedure Difference in Your LaboratoryNo
QuantityAmount-of-substance concentration
From (0)0.5
To (0)25
Unit (0)mmol/l
From (1)1
To (1)1
Unit (1)%
Level of Confidence (%)95%
Comments on Concentration RangeDetermination of higher concentrations after predilution of serum
Reference Procedure Status AcceptableYes
Laboratory StatusB
Survey OrganisationIFCC - EQAS
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Participation Status in Intercomparison Surveys acceptableYes
Recommended for Publication in the JCTLM ListYes
Organization AcronymRfB
Organization NameReference Institute for Bioanalytics, Calibration Laboratory II
Organization Websitehttps://rfb.bio/cgi/switchLang?lang=en
Organization Country ISO CodeDE
Organization Country NameGermany
Analyte Category NameMetabolites and substrates