Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C1RMS146 - alanine aminotransferase (ALT) in Blood serum

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC1RMS146
Reference Laboratory StatusCalibration laboratory of the DGKL
Analyte Namealanine aminotransferase (ALT)
IUPAC IFCC NumberEC 2.6.1.2
Traceable to SIYes
Principle of Analytical ProcedureKinetic spectrophotometry
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureIFCC Reference Measurement Procedure (37°C) for ALT
JCTLM Database Identifier NumberNRMeth 67
QuantityCatalytic activity concentration
From (0)0.072
To (0)4.75
Unit (0)µkat/l
From (1)2.2
To (1)2.2
Unit (1)%
Level of Confidence (%)95%
Comments on Concentration RangeThe measurement range spans from 0.072 µkat/l (4.3 U/l) to 4.75 µkat/l (285 U/l) without dilution of the sample
Reference Procedure Status AcceptableUncertainty appears wrongly indicated > instead of <.
Laboratory Preparing for AccreditationYes
Survey OrganisationIFCC - EQAS
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Participation Status in Intercomparison Surveys acceptableYES
Recommended for Publication in the JCTLM ListYes
Organization AcronymRfB
Organization NameReference Institute for Bioanalytics, Calibration Laboratory II
Organization Websitehttps://rfb.bio/cgi/switchLang?lang=en
Organization Country ISO CodeDE
Organization Country NameGermany
Analyte Category NameEnzymes