JCTLM Database: higher-order reference materials, methods and services
Details : C1RMS25 - glucose in Whole blood, Blood serum, Blood plasma, Other, Calibration solution
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C1RMS25 |
| Reference Laboratory Status | accredited according ISO 17025 |
| Analyte Name | glucose |
| IUPAC IFCC Number | CAS50-99-7 |
| Traceable to SI | yes |
| Principle of Analytical Procedure | Isotope dilution gas chromatography mass spectrometry (ID/GC/MS) |
| Reference Method Listed in JCTLM | yes |
| Name of JCTLM Listed Reference Method Procedure | NIST definitive method for Serum Glucose |
| JCTLM Database Identifier Number | NRMeth 80 |
| Procedure Difference in Your Laboratory | no |
| Quantity | Mass concentration |
| From (0) | 12 |
| To (0) | 800 |
| Unit (0) | mg/dl |
| From (1) | 1.7 |
| To (1) | 1.7 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Comments on Uncertainty Range | The expanded uncertainty calculated according to GUM |
| Reference Procedure Status Acceptable | Yes |
| Laboratory Preparing for Accreditation | n.a. |
| Laboratory Status | B |
| Survey Organisation | DGKL, IFCC, Instand |
| Intercomparison Results Publicly Available | yes |
| Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
| Participation Status in Intercomparison Surveys acceptable | Yes |
| Recommended for Publication in the JCTLM List | Yes |
| Organization Acronym | RDG |
| Organization Name | Roche Diagnostics GmbH |
| Organization Website | http://www.roche.com |
| Organization Country ISO Code | DE |
| Organization Country Name | Germany |
| Analyte Category Name | Metabolites and substrates |