Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C1RMS25 - glucose in Whole blood, Blood serum, Blood plasma, Other, Calibration solution

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC1RMS25
Reference Laboratory Statusaccredited according ISO 17025
Analyte Nameglucose
IUPAC IFCC NumberCAS50-99-7
Traceable to SIyes
Principle of Analytical ProcedureIsotope dilution gas chromatography mass spectrometry (ID/GC/MS)
Reference Method Listed in JCTLMyes
Name of JCTLM Listed Reference Method ProcedureNIST definitive method for Serum Glucose
JCTLM Database Identifier NumberNRMeth 80
Procedure Difference in Your Laboratoryno
QuantityMass concentration
From (0)12
To (0)800
Unit (0)mg/dl
From (1)1.7
To (1)1.7
Unit (1)%
Level of Confidence (%)95%
Comments on Uncertainty RangeThe expanded uncertainty calculated according to GUM
Reference Procedure Status AcceptableYes
Laboratory Preparing for Accreditationn.a.
Laboratory StatusB
Survey OrganisationDGKL, IFCC, Instand
Intercomparison Results Publicly Availableyes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Participation Status in Intercomparison Surveys acceptableYes
Recommended for Publication in the JCTLM ListYes
Organization AcronymRDG
Organization NameRoche Diagnostics GmbH
Organization Websitehttp://www.roche.com
Organization Country ISO CodeDE
Organization Country NameGermany
Analyte Category NameMetabolites and substrates