Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C1RMS36 - aldosterone in Blood serum, Blood plasma, Calibration solution

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC1RMS36
Reference Laboratory StatusEQAS Organizer, University Hospital
Analyte Namealdosterone
IUPAC IFCC NumberCAS52-39-1
Applicable MatricesBlood serum, Blood plasma, Calibration solution
Traceable to SIYes
Principle of Analytical ProcedureIsotope dilution mass spectrometry (IDMS) SPMD-RfB currently provides the service for aldosterone using a reference measurement procedure based on LC-MS/MS principle.
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureDGKC definitive Method for Serum Aldosterone
JCTLM Database Identifier NumberNRMeth 56
Procedure Difference in Your LaboratoryNo
QuantityAmount-of-substance concentration
From (0)0.05
To (0)3.5
Unit (0)nmol/l
From (1)1
To (1)1
Unit (1)%
Level of Confidence (%)95%
Reference Procedure Status AcceptableYes
Laboratory Statusb
Survey OrganisationIFCC-EQAS for Reference Laboratories
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Participation Status in Intercomparison Surveys acceptableYes
Comments for JCTLM DB PublicationSPMD-RfB currently provides the service for aldosterone using a reference measurement procedure based on LC-MS/MS principle.
Recommended for Publication in the JCTLM ListYes
Organization AcronymSPMD-RfB
Organization NameReference Institute for Bioanalysis, Calibration Laboratory I
Organization Websitehttps://www.rfb.bio
Organization Country ISO CodeDE
Organization Country NameGermany
Analyte Category NameNon-peptide hormones