Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C1RMS44 - digoxin in Blood serum, Blood plasma, Calibration solution

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC1RMS44
Reference Laboratory StatusEQAS Organizer, University Hospital
Analyte Namedigoxin
IUPAC IFCC NumberCAS20830-75-5
Traceable to SIYes
Principle of Analytical ProcedureIsotope dilution mass spectrometry (IDMS)
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureDGKC definitive Method for Serum Digoxine
JCTLM Database Identifier NumberNRMeth 54
Procedure Difference in Your LaboratoryNo
QuantityAmount-of-substance concentration
From (0)0.1
To (0)40
Unit (0)nmol/l
From (1)1
To (1)1
Unit (1)%
Level of Confidence (%)95%
Reference Procedure Status AcceptableYes
Laboratory Status(b) Accredited (ISO 17025+15195)
Survey OrganisationIFCC-EQAS for Reference Laboratories
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Participation Status in Intercomparison Surveys acceptableYes
Recommended for Publication in the JCTLM ListYes
Organization AcronymSPMD-RfB
Organization NameReference Institute for Bioanalysis, Calibration Laboratory I
Organization Websitehttps://www.rfb.bio
Organization Country ISO CodeDE
Organization Country NameGermany
Analyte Category NameDrugs