JCTLM Database: higher-order reference materials, methods and services
Details : C1RMS72 - lithium in Blood serum
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C1RMS72 |
Reference Laboratory Status | EQAS Organiser |
Analyte Name | lithium |
IUPAC IFCC Number | CAS7439-93-2 |
Applicable Matrices | serum |
Traceable to SI | Yes |
Principle of Analytical Procedure | Atomic absorption spectrometry |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | NIST/CDC/AACC Flame Atomic Absorbtion Method for Serum Lithium |
JCTLM Database Identifier Number | C1RMP_El1 |
Procedure Difference in Your Laboratory | N |
Quantity | Amount-of-substance concentration |
From (0) | 0.2 |
To (0) | 2 |
Unit (0) | mmol/L |
From (1) | 1.91 |
To (1) | 1.91 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Uncertainty Range | Coverage factor of 2. The expanded uncertainty calculated according to GUM |
Laboratory Preparing for Accreditation | Yes |
Intercomparison Results Publicly Available | Yes |
Recommended for Publication in the JCTLM List | Yes |
RMS Provider Comment in Case of Delisting | Service re-instated on 07 August 2012 |
Organization Acronym | WEQAS |
Organization Name | Quality Laboratories, WEQAS |
Organization Website | http://www.weqas.com |
Organization Country ISO Code | GB |
Organization Country Name | United Kingdom |
Analyte Category Name | Electrolytes |