Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C1RMS97C - glycated hemoglobin (HbA1c) in Whole blood

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC1RMS97C
Reference Laboratory StatusEQAS-Organiser
Analyte Nameglycated hemoglobin (HbA1c)
IUPAC IFCC NumberQU60433
Applicable Matricesfresh whole blood, frozen or lyophilized haemolysates
Traceable to SIYes
Principle of Analytical ProcedureLC-MS
Reference Method Listed in JCTLMYes(Jeppson) No (Kaiser)
Name of JCTLM Listed Reference Method ProcedureIFCC Method for HbA1c
JCTLM Database Identifier NumberC1RMP_P16_MS
Procedure Difference in Your LaboratoryYes
QuantityAmount-of-substance fraction
From (0)29
To (0)150
Unit (0)mmol/mol
From (1)1.5
To (1)1.5
Unit (1)%
Level of Confidence (%)95%
Comments on Uncertainty RangeThe stated expanded uncertainty value corresponds to the best measurement capability.
Reference Procedure Status AcceptableYes
Laboratory Preparing for AccreditationYes
Laboratory StatusYes
Survey OrganisationIFCC (RELA) IFCC-HbA1c-Network
Intercomparison Results Publicly AvailableYes (RELA) No(HbA1c-Network)
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Reference to Published Comparison ResultsData available on request from Instand e.V. reference laboratory
Participation Status in Intercomparison Surveys acceptableYes
Comments of Review TeamsRMP under review in JCTLM WG1 Cycle IV and accepted
Recommended for Publication in the JCTLM ListYes
Organization AcronymINSTAND e. V.
Organization NameSociety for Promoting Quality Assurance in Medical Laboratories
Organization Websitehttp://www.instand-ev.de
Organization Country ISO CodeDE
Organization Country NameGermany
Analyte Category NameProteins