Details : C20RMP4 - GPHCM reference measurement method/procedure for aldosterone in human plasma

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC20RMP4
Analyte Namealdosterone
Method capable of Traceability of Analyte to SI or Defined Procedure
SI
Applicable Matriceshuman plasma; fresh, frozen or lyophilized
NameGPHCM reference measurement method/procedure for aldosterone in human plasma
Organization that developed Validated MethodGuangdong Provincial Hospital of Chinese Medicine (GPHCM)
Measurement Techniques UsedIsotope dilution liquid chromatography mass spectrometry
QuantityAmount-of-substance concentration
From (0)0.0277
To (0)2.7744
Unit (0)nmol/L
From (1)0.009
To (1)0.039
Unit (1)nmol/L
Level of Confidence (%)95%
Peer Reviewed PublicationAn isotope dilution liquid chromatography-tandem mass spectrometry candidate reference measurement procedure for aldosterone measurement in human plasma. Zhuang J. et al, Analytical and Bioanalytical Chemistry 2021 413 4471-4481
Hyperlink to Final Report of extent of Equivalence Studies amongC20RMP4 Extent-of-equivalence report for CGPHM RMP ALD.pdf
Analyte Category NameNon-peptide hormones
Non-CCQM Key Comparison ReportC20 RMP4 Interlaboratory comparison report for ALD RMP
Non-CCQM Key Comparison Report Hyperlinkhttps://www.jctlmdb.org/backend/api/jctlmFile/downloadFile/825/16
Measurement principle/techniqueIsotope dilution mass spectrometry
Measurement principle/techniqueLiquid chromatography mass spectrometry