Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C20RMP8 - ID-LC-MS/MS-based reference measurement procedure for the quantification of carbamazepine-10,11-epoxide in human serum and plasma

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC20RMP8
Analyte Namecarbamazepine 10, 11-epoxide
IUPAC IFCC NumberCAS36507-30-9
Method capable of Traceability of Analyte to SI or Defined Procedureyes
Applicable Matriceshuman serum and plasma; fresh or frozen
NameID-LC-MS/MS-based reference measurement procedure for the quantification of carbamazepine-10,11-epoxide in human serum and plasma
Organization that developed Validated MethodThis reference measurement procedure was developed and validated by the laboratory of Dr. Risch Ostschweiz AG in Buchs SG, Switzerland in collaboration with Roche Diagnostics GmbH in Penzberg, Germany
Measurement Techniques UsedIsotope dilution mass spectrometry
QuantityMass concentration
From (0)0.04
To (0)12
Unit (0)µg/mL
From (1)4.3
To (1)3.2
Unit (1)%
Level of Confidence (%)0.95
Comments on Uncertainty RangeThe expected measurement uncertainty of 4.3 % is at the mass concentration of 0.120 µg/mL. The expected measurement uncertainty of 3.2 % is at the mass concentration of 6.00 µg/mL.
Peer Reviewed PublicationAn isotope dilution-liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS)-based candidate reference measurement procedure (RMP) for the quantification of carbamazepine 10, 11-epoxide in human serum and plasma, Taibon J. et al, Clinical Chemisty and Laboratory Medicine 2024 62(7) 1339-1351
Comments to be published via JCTLM DBThis reference measurement procedure was developed and validated by the laboratory of Dr. Risch Ostschweiz AG in Buchs SG, Switzerland in collaboration with Roche Diagnostics GmbH in Penzberg, Germany
Analyte Category NameDrugs
Non-CCQM Key Comparison ReportSee CCLM publication cited above for comparability assessment study
Non-CCQM Key Comparison Report Hyperlinknot applicable
Measurement principle/techniqueIsotope dilution mass spectrometry