JCTLM Database: higher-order reference materials, methods and services
Details : C3RMMP11 - NIST LC/MS/MS method for norandrosterone
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | C3RMMP11 |
| Analyte Name | norandrosterone |
| IUPAC IFCC Number | norandrosterone |
| Method capable of Traceability of Analyte to SI or Defined Procedure | yes |
| Applicable Matrices | fresh or frozen human urine |
| Name | NIST LC/MS/MS method for norandrosterone |
| Organization that developed Validated Method | NIST, Gaithersburg, MD USA |
| Measurement Techniques Used | ID/LC/MS/MS |
| Quantity | Mass concentration |
| From (0) | 1 |
| To (0) | 10 |
| Unit (0) | ng/mL |
| From (1) | 4.6 |
| To (1) | 4.1 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Peer Reviewed Publication | Anal. Chem., 2006, 78, 3393-3398
|
| Measurand Clearly Defined | yes |
| Multiple Testing Sites Not Required to obtain a Single Value | correct |
| Analyte Category Name | Non-peptide hormones |
| Measurement principle/technique | Isotope dilution mass spectrometry |