JCTLM Database: higher-order reference materials, methods and services
Details : C3RMMP11 - NIST LC/MS/MS method for norandrosterone
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | C3RMMP11 |
Analyte Name | norandrosterone |
IUPAC IFCC Number | norandrosterone |
Method capable of Traceability of Analyte to SI or Defined Procedure | yes |
Applicable Matrices | fresh or frozen human urine |
Name | NIST LC/MS/MS method for norandrosterone |
Organization that developed Validated Method | NIST, Gaithersburg, MD USA |
Measurement Techniques Used | ID/LC/MS/MS |
Quantity | Mass concentration |
From (0) | 1 |
To (0) | 10 |
Unit (0) | ng/mL |
From (1) | 4.6 |
To (1) | 4.1 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | Anal. Chem., 2006, 78, 3393-3398
|
Measurand Clearly Defined | yes |
Multiple Testing Sites Not Required to obtain a Single Value | correct |
Analyte Category Name | Non-peptide hormones |
Measurement principle/technique | Isotope dilution mass spectrometry |