JCTLM Database: higher-order reference materials, methods and services
Details : C3RMMP19 - DGKL reference method for glucose
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | C3RMMP19 |
| Analyte Name | glucose |
| IUPAC IFCC Number | CAS50-99-7 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | lyophilized, fresh or frozen blood serum, blood plasma, whole blood, urine, or calibration solution; and, other biological material |
| Name | DGKL reference method for glucose |
| Organization that developed Validated Method | German Society of Clinical Chemistry and Laboratory Medicine (DGKL) |
| Measurement Techniques Used | ID/GC/MS |
| Comments 1 | Calibrated with commercial reference material from NIST |
| Quantity | Amount-of-substance-concentration |
| From (0) | 0.5 |
| To (0) | 20 |
| Unit (0) | mmol/l |
| Level of Confidence (%) | 95% |
| Peer Reviewed Publication | Biomedical Mass Spectrometry, 1982, 9, 395-405
|
| How to Obtain Method Copy | schumann.gerhard@mh-hannover.de |
| Multiple Testing Sites Not Required to obtain a Single Value | No |
| Analyte Category Name | Metabolites and substrates |
| Measurement principle/technique | Isotope dilution mass spectrometry |