JCTLM Database: higher-order reference materials, methods and services
Details : C3RMMP20 - DGKL reference method for theophylline
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | C3RMMP20 |
| Analyte Name | theophylline |
| IUPAC IFCC Number | CAS58-55-9 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | Yes |
| Applicable Matrices | lyophilized, fresh or frozen blood serum, blood plasma, or calibration solution; and, other biological material |
| Name | DGKL reference method for theophylline |
| Organization that developed Validated Method | German Society of Clinical Chemistry and Laboratory Medicine (DGKL) |
| Measurement Techniques Used | ID/GC/MS |
| Comments 1 | Calibrated with commercial standard material |
| Quantity | Mass concentration |
| From (0) | 4 |
| To (0) | 20 |
| Unit (0) | mg/l |
| Level of Confidence (%) | 95% |
| Peer Reviewed Publication | Fresenius Z. Anal. Chem., 1984, 317, 677
J. Chrom., 1984, 336, 285-291
|
| How to Obtain Method Copy | schumann.gerhard@mh-hannover.de |
| Multiple Testing Sites Not Required to obtain a Single Value | No |
| Review Team Name | Drugs |
| Analyte Category Name | Drugs |
| Measurement principle/technique | Isotope dilution mass spectrometry |