Details : C3RMMP21 - DGKL reference method for ortho-phosphate
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | C3RMMP21 |
| Analyte Name | ortho-phosphate |
| IUPAC IFCC Number | CAS7664-38-2 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | lyophilized, fresh or frozen blood serum, calibration solution, or urine; and other biological material |
| Name | DGKL reference method for ortho-phosphate |
| Organization that developed Validated Method | German Society of Clinical Chemistry and Laboratory Medicine (DGKL) |
| Measurement Techniques Used | HPIC (conductivity) |
| Comments 1 | Calibrated with commercial reference material from NIST |
| Quantity | Amount-of-substance-concentration |
| From (0) | 0.8 |
| To (0) | 26 |
| Unit (0) | mmol/l |
| From (1) | 2 |
| Unit (1) | % relative |
| Level of Confidence (%) | 95% |
| Peer Reviewed Publication | Fresenius J. Anal. Chem., 1990, 337, 143
|
| Multiple Testing Sites Not Required to obtain a Single Value | No |
| Analyte Category Name | Electrolytes |
| Measurement principle/technique | Ion chromatography |