Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C3RMMP21 - DGKL reference method for ortho-phosphate

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC3RMMP21
Analyte Nameortho-phosphate
IUPAC IFCC NumberCAS7664-38-2
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matriceslyophilized, fresh or frozen blood serum, calibration solution, or urine; and other biological material
NameDGKL reference method for ortho-phosphate
Organization that developed Validated MethodGerman Society of Clinical Chemistry and Laboratory Medicine (DGKL)
Measurement Techniques UsedHPIC (conductivity)
Comments 1Calibrated with commercial reference material from NIST
QuantityAmount-of-substance-concentration
From (0)0.8
To (0)26
Unit (0)mmol/l
From (1)2
Unit (1)% relative
Level of Confidence (%)95%
Peer Reviewed PublicationFresenius J. Anal. Chem., 1990, 337, 143
Multiple Testing Sites Not Required to obtain a Single ValueNo
Analyte Category NameElectrolytes
Measurement principle/techniqueIon chromatography