JCTLM Database: higher-order reference materials, methods and services
Details : C3RMMP21 - DGKL reference method for ortho-phosphate
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | C3RMMP21 |
Analyte Name | ortho-phosphate |
IUPAC IFCC Number | CAS7664-38-2 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | lyophilized, fresh or frozen blood serum, calibration solution, or urine; and other biological material |
Name | DGKL reference method for ortho-phosphate |
Organization that developed Validated Method | German Society of Clinical Chemistry and Laboratory Medicine (DGKL) |
Measurement Techniques Used | HPIC (conductivity) |
Comments 1 | Calibrated with commercial reference material from NIST |
Quantity | Amount-of-substance-concentration |
From (0) | 0.8 |
To (0) | 26 |
Unit (0) | mmol/l |
From (1) | 2 |
Unit (1) | % relative |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | Fresenius J. Anal. Chem., 1990, 337, 143
|
Multiple Testing Sites Not Required to obtain a Single Value | No |
Analyte Category Name | Electrolytes |
Measurement principle/technique | Ion chromatography |