JCTLM Database: higher-order reference materials, methods and services

Details : C3RMMP35 - PTB reference method for blood cell counting, thrombocyte

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC3RMMP35
Analyte Namethrombocyte
IUPAC IFCC Numberthrombocyte
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matricesfresh whole blood
NamePTB reference method for blood cell counting, thrombocyte
Organization that developed Validated MethodPhysikalisch-Technische Bundesanstalt (PTB)
Measurement Techniques UsedFlow cytometry, cell differentiation by fluorescence detection of antibody stained platelets and light scattering, coincidence correction by dilution series
QuantityParticle concentration
From (0)30
To (0)1500
Unit (0)(1/nl)
From (1)10
To (1)10
Unit (1)% relative
Level of Confidence (%)95%
Peer Reviewed PublicationJ. Lab. Med., 2012, 36(1), 25-35 German National Standard DIN 58932-5
How to Obtain Method CopyContact DIN
Measurand Clearly Definedyes
Multiple Testing Sites Not Required to obtain a Single ValueMultiple testing sites not required
Additional CommentsValidation of the reference method described in DIN 58932-5 is presently accomplished
Comments to be published via JCTLM DBValidation of the reference method described in DIN 58932-5 is presently accomplished. Procedures used to derive the thrombocyte concentration by determining the platelet / erythrocyte ratio (Am. J. Clin. Pathol., 2001, 115, 480-484 and references therein) are not suited as primary reference procedure, since coincidence correction has not been proved to yield correct results for particle subpopulations exhibiting dramatically different concentrations (e.g. 10 platelets / nl compared to 4000 erythrocytes / nl)
Analyte Category NameBlood cell counting
Measurement principle/techniqueFlow cytometry