JCTLM Database: higher-order reference materials, methods and services
Details : C3RMMP5 - NIST LC/MS/MS reference method for folates (folic acid)
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | C3RMMP5 |
Analyte Name | folic acid |
IUPAC IFCC Number | CAS59-30-3 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | human serum |
Name | NIST LC/MS/MS reference method for folates (folic acid) |
Organization that developed Validated Method | NIST |
Measurement Techniques Used | LC/MS/MS |
Comments 1 | Isotope Dilution |
Quantity | Mass concentration |
From (0) | 2 |
To (0) | 25 |
Unit (0) | ng/mL |
From (1) | 2 |
To (1) | 6 |
Unit (1) | % |
Peer Reviewed Publication | Simultaneous quantification of homocysteine and folate in human serum or plasma using liquid chromatography tandem mass spectrometry, Nelson B C, Anal. Chem., 2005, 77(11), 3586-3593
|
Measurand Clearly Defined | yes |
Review Team Name | vitamins |
Analyte Category Name | Vitamins and micronutrients |
Non-CCQM Key Comparison Report | SRM 1955 value assignment manuscript - Anal. Bioanal. Chem., 2006, 385, 612-622
|
Measurement principle/technique | Liquid chromatography mass spectrometry |