JCTLM Database: higher-order reference materials, methods and services
Details : C4RMP1 - NIST ID-LC/MS method for serum creatinine
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | C4RMP1 |
Analyte Name | creatinine |
IUPAC IFCC Number | CAS60-27-5 |
Method capable of Traceability of Analyte to SI or Defined Procedure | Yes |
Applicable Matrices | fresh, frozen, and lyophilized human serum |
Name | NIST ID-LC/MS method for serum creatinine |
Organization that developed Validated Method | NIST |
Measurement Techniques Used | ID-LC/MS |
Quantity | Amount-of-substance concentration |
From (0) | 6 |
To (0) | 60 |
Unit (0) | mg/L |
From (1) | 2 |
To (1) | 1 |
Unit (1) | % relative |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | Clin. Chem., 2007, 53:9, 1694-1699
|
How to Obtain Method Copy | nathan.dodder@nist.gov |
Measurand Clearly Defined | Yes |
Multiple Testing Sites Not Required to obtain a Single Value | yes |
Review Team Name | metabolites and substrates |
Analyte Category Name | Metabolites and substrates |
Measurement principle/technique | Isotope dilution mass spectrometry |