Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C4RMP1 - NIST ID-LC/MS method for serum creatinine

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC4RMP1
Analyte Namecreatinine
IUPAC IFCC NumberCAS60-27-5
Method capable of Traceability of Analyte to SI or Defined ProcedureYes
Applicable Matricesfresh, frozen, and lyophilized human serum
NameNIST ID-LC/MS method for serum creatinine
Organization that developed Validated MethodNIST
Measurement Techniques UsedID-LC/MS
QuantityAmount-of-substance concentration
From (0)6
To (0)60
Unit (0)mg/L
From (1)2
To (1)1
Unit (1)% relative
Level of Confidence (%)95%
Peer Reviewed PublicationClin. Chem., 2007, 53:9, 1694-1699
How to Obtain Method Copynathan.dodder@nist.gov
Measurand Clearly DefinedYes
Multiple Testing Sites Not Required to obtain a Single Valueyes
Review Team Namemetabolites and substrates
Analyte Category NameMetabolites and substrates
Measurement principle/techniqueIsotope dilution mass spectrometry