JCTLM Database: higher-order reference materials, methods and services
Details : C4RMP2 - NIST ID-LC/MS method for serum testosterone
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | C4RMP2 |
| Analyte Name | testosterone |
| IUPAC IFCC Number | CAS58-22-0 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | Yes |
| Applicable Matrices | fresh and frozen human serum |
| Name | NIST ID-LC/MS method for serum testosterone |
| Organization that developed Validated Method | NIST |
| Measurement Techniques Used | ID-LC/MS |
| Quantity | Amount-of-substance concentration |
| From (0) | 0.3 |
| To (0) | 8.5 |
| Unit (0) | ng/g |
| From (1) | 2.5 |
| To (1) | 2.5 |
| Unit (1) | % relative |
| Level of Confidence (%) | 95% |
| Peer Reviewed Publication | Manuscript submitted to Analytical and Bioanalytical Chemistry |
| How to Obtain Method Copy | susan.tai@nist.gov |
| Measurand Clearly Defined | Yes |
| Multiple Testing Sites Not Required to obtain a Single Value | yes |
| Review Team Name | non-peptide hormones |
| Analyte Category Name | Non-peptide hormones |
| Measurement principle/technique | Isotope dilution mass spectrometry |