Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C4RMS3 - total cholesterol in Blood serum

Type (acronym)rms
Typereference measurement service
Unique Nomination NumberC4RMS3
Reference Laboratory StatusEQAS Organiser
Analyte Nametotal cholesterol
IUPAC IFCC Numbertotal cholesterol
Applicable Matricesserum
Traceable to SIYes
Principle of Analytical ProcedureID-GCMS (Isotope Dilution Gas Chromatography Mass Spectrometry)
Reference Method Listed in JCTLMYes
Name of JCTLM Listed Reference Method ProcedureUniversity of Ghent reference method for cholesterol
JCTLM Database Identifier NumberNRMeth 2
Procedure Difference in Your LaboratoryNo
QuantityAmount-of-substance concentration
From (0)1
To (0)8
Unit (0)mmol/L
From (1)3.63
To (1)3.63
Unit (1)%
Level of Confidence (%)95%
Laboratory Preparing for AccreditationPre-assessment visit July 2008, full assessment due April 2009
Survey OrganisationJCTLM RELA2007
Intercomparison Results Publicly AvailableYes
Link to Comparison ResultsRELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml
Additional Information 1Cholesterol RELA2007 results removed from database as incorrect results submitted initially. Final submitted results agreed with peer group, although these are not displayed on the website as they were submitted late.
Recommended for Publication in the JCTLM ListYes
RMS Provider Comment in Case of DelistingService re-instated on 07 August 2012
Organization AcronymWEQAS
Organization NameQuality Laboratories, WEQAS
Organization Websitehttp://www.weqas.com
Organization Country ISO CodeGB
Organization Country NameUnited Kingdom
Analyte Category NameMetabolites and substrates