JCTLM Database: higher-order reference materials, methods and services
Details : C6RMS4 - total glycerides in Blood serum
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C6RMS4 |
Reference Laboratory Status | EQAS Organiser |
Analyte Name | total glycerides |
IUPAC IFCC Number | total glycerides |
Applicable Matrices | serum |
Traceable to SI | Yes |
Principle of Analytical Procedure | ID-GCMS |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | University of Ghent reference measurement procedure for triglycerides |
JCTLM Database Identifier Number | NRMeth 5 |
Procedure Difference in Your Laboratory | No |
Quantity | Amount-of-substance concentration |
From (0) | 0.6 |
To (0) | 8 |
Unit (0) | mmol/L |
From (1) | 2.83 |
To (1) | 2.83 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Survey Organisation | JCTLM RELA2009 |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Recommended for Publication in the JCTLM List | Yes |
Organization Acronym | WEQAS |
Organization Name | Quality Laboratories, WEQAS |
Organization Website | http://www.weqas.com |
Organization Country ISO Code | GB |
Organization Country Name | United Kingdom |
Analyte Category Name | Metabolites and substrates |