JCTLM Database: higher-order reference materials, methods and services
Details : C7 RMS1 - theophylline in Blood serum, Blood plasma
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C7 RMS1 |
| Reference Laboratory Status | EQAS Organizer |
| Analyte Name | theophylline |
| IUPAC IFCC Number | CAS58-55-9 |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | GC-IDMS |
| Reference Method Listed in JCTLM | Yes |
| Name of JCTLM Listed Reference Method Procedure | INSTAND Reference Measurement Procedure for Theophylline |
| JCTLM Database Identifier Number | C4RMP6 |
| Procedure Difference in Your Laboratory | No |
| Quantity | Amount-of-substance concentration |
| From (0) | 5 |
| To (0) | 500 |
| Unit (0) | µmol/L |
| From (1) | 1 |
| To (1) | 1 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Reference Procedure Status Acceptable | Yes (procedure listed) |
| Laboratory Status | (b) accredited |
| Survey Organisation | IFCC (Rela) |
| Intercomparison Results Publicly Available | Yes |
| Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
| Participation Status in Intercomparison Surveys acceptable | yes |
| Recommended for Publication in the JCTLM List | Yes |
| Organization Acronym | INSTAND e. V. |
| Organization Name | Society for Promoting Quality Assurance in Medical Laboratories |
| Organization Website | http://www.instand-ev.de |
| Organization Country ISO Code | DE |
| Organization Country Name | Germany |
| Analyte Category Name | Drugs |