JCTLM Database: higher-order reference materials, methods and services
Details : C7 RMS2 - urea in Blood serum, Blood plasma
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C7 RMS2 |
| Reference Laboratory Status | EQAS Organizer |
| Analyte Name | urea |
| IUPAC IFCC Number | CAS57-13-6 |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | GC-IDMS |
| Reference Method Listed in JCTLM | Yes |
| Name of JCTLM Listed Reference Method Procedure | DGKC Reference Method for Urea |
| JCTLM Database Identifier Number | NRMeth 52 |
| Procedure Difference in Your Laboratory | Yes |
| Quantity | Amount-of-substance concentration |
| From (0) | 0.5 |
| To (0) | 50 |
| Unit (0) | mmol/L |
| From (1) | 1 |
| To (1) | 1 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Survey Organisation | IFCC (Rela) |
| Intercomparison Results Publicly Available | Yes |
| Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
| Recommended for Publication in the JCTLM List | Yes |
| Organization Acronym | INSTAND e. V. |
| Organization Name | Society for Promoting Quality Assurance in Medical Laboratories |
| Organization Website | http://www.instand-ev.de |
| Organization Country ISO Code | DE |
| Organization Country Name | Germany |
| Analyte Category Name | Metabolites and substrates |