JCTLM Database: higher-order reference materials, methods and services
Details : C7 RMS3 - glucose in Blood serum, Calibration solution
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C7 RMS3 |
Reference Laboratory Status | National Metrology Institute |
Analyte Name | glucose |
IUPAC IFCC Number | CAS50-99-7 |
Applicable Matrices | lyophilized, fresh, or frozen human serum, calibration solution |
Traceable to SI | Yes |
Principle of Analytical Procedure | ID-GC/MS |
Reference Method Listed in JCTLM | No |
Name of JCTLM Listed Reference Method Procedure | University of Ghent reference method for glucose |
JCTLM Database Identifier Number | NRMeth 4 |
Procedure Difference in Your Laboratory | YES |
Quantity | Amount-of-substance concentration |
From (0) | 1.6 |
To (0) | 20 |
Unit (0) | mmol/L |
From (1) | 2 |
To (1) | 1 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Uncertainty Range | The expanded uncertainty is relative. |
Survey Organisation | IFCC RELA 2010 (Lab 105) |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Additional Information 1 | LNE will participate in CCQM-K11.2 in 2012 |
Recommended for Publication in the JCTLM List | Yes |
Organization Acronym | LNE |
Organization Name | Laboratoire National de Métrologie et d'Essais |
Organization Website | http://www.lne.fr |
Organization Country ISO Code | FR |
Organization Country Name | France |
Analyte Category Name | Metabolites and substrates |