JCTLM Database: higher-order reference materials, methods and services
Details : C7 RMS5 - total cholesterol in Blood serum, Calibration solution
| Type (acronym) | rms |
| Type | reference measurement service |
| Unique Nomination Number | C7 RMS5 |
| Reference Laboratory Status | National Metrology Institute |
| Analyte Name | total cholesterol |
| IUPAC IFCC Number | total cholesterol |
| Applicable Matrices | lyophilized, fresh, or frozen human serum, calibration solution |
| Traceable to SI | Yes |
| Principle of Analytical Procedure | ID-GC/MS |
| Reference Method Listed in JCTLM | No |
| Name of JCTLM Listed Reference Method Procedure | DGKC reference method for serum cholesterol |
| JCTLM Database Identifier Number | NRMeth 46 |
| Procedure Difference in Your Laboratory | YES |
| Quantity | Amount-of-substance concentration |
| From (0) | 1 |
| To (0) | 10 |
| Unit (0) | mmol/L |
| From (1) | 3 |
| To (1) | 1 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Comments on Uncertainty Range | The expanded uncertainty is relative. |
| Survey Organisation | IFCC RELA 2011 (Lab 105) |
| Intercomparison Results Publicly Available | Yes |
| Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
| Additional Information 1 | LNE will participate in CCQM-K6.3 in 2012 |
| Recommended for Publication in the JCTLM List | Yes |
| Organization Acronym | LNE |
| Organization Name | Laboratoire National de Métrologie et d'Essais |
| Organization Website | http://www.lne.fr |
| Organization Country ISO Code | FR |
| Organization Country Name | France |
| Analyte Category Name | Metabolites and substrates |