JCTLM Database: higher-order reference materials, methods and services
Details : C7RMP17 - NIST method for 25-hydroxyvitamin D3
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | C7RMP17 |
| Analyte Name | 25-hydroxyvitamin D3 |
| IUPAC IFCC Number | CAS63283-36-3 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | serum, plasma |
| Name | NIST method for 25-hydroxyvitamin D3 |
| Organization that developed Validated Method | NIST |
| Measurement Techniques Used | ID-LC-MS/MS |
| Quantity | Mass fraction |
| From (0) | 6.311 |
| To (0) | 26.972 |
| Unit (0) | ng/g |
| From (1) | 2 |
| To (1) | 2.1 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Peer Reviewed Publication | Anal. Chem., 2010, 82, 1942
|
| Analyte Category Name | Metabolites and substrates |
| Measurement principle/technique | Isotope dilution mass spectrometry |