Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C7RMP26 - NIM reference method for serum creatinine

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC7RMP26
Analyte Namecreatinine
IUPAC IFCC NumberCAS60-27-5
Method capable of Traceability of Analyte to SI or Defined ProcedureSI
Applicable Matriceslyophilized, fresh, or frozen human serum
NameNIM reference method for serum creatinine
Organization that developed Validated MethodNIM
Measurement Techniques UsedID/LC/MS, ID/LC/MS/MS
Comments 2Calibrate with SRM 914a Creatinine
QuantityMass fraction
From (0)2
To (0)50
Unit (0)µg/g
From (1)0.5
To (1)2
Unit (1)%
Level of Confidence (%)95%
Peer Reviewed PublicationAnal. Let., 2008, 41, 2912-2922
Additional CommentsThere is no writing document available for the method that is totally compliant with the mandatory elements of ISO 15193.
Review Team NameMetabolites and substrates
Analyte Category NameMetabolites and substrates
Measurement principle/techniqueIsotope dilution mass spectrometry