JCTLM Database: higher-order reference materials, methods and services
Details : C7RMP26 - NIM reference method for serum creatinine
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | C7RMP26 |
Analyte Name | creatinine |
IUPAC IFCC Number | CAS60-27-5 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | lyophilized, fresh, or frozen human serum |
Name | NIM reference method for serum creatinine |
Organization that developed Validated Method | NIM |
Measurement Techniques Used | ID/LC/MS, ID/LC/MS/MS |
Comments 2 | Calibrate with SRM 914a Creatinine |
Quantity | Mass fraction |
From (0) | 2 |
To (0) | 50 |
Unit (0) | µg/g |
From (1) | 0.5 |
To (1) | 2 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | Anal. Let., 2008, 41, 2912-2922
|
Additional Comments | There is no writing document available for the method that is totally compliant with the mandatory elements of ISO 15193. |
Review Team Name | Metabolites and substrates |
Analyte Category Name | Metabolites and substrates |
Measurement principle/technique | Isotope dilution mass spectrometry |