JCTLM Database: higher-order reference materials, methods and services
Details : C7RMP27 - NIM reference method for serum uric acid
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | C7RMP27 |
| Analyte Name | uric acid |
| IUPAC IFCC Number | CAS69-93-2 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | lyophilized, fresh, or frozen human serum |
| Name | NIM reference method for serum uric acid |
| Organization that developed Validated Method | NIM |
| Measurement Techniques Used | ID/LC/MS, ID/LC/MS/MS |
| Comments 2 | Calibrate with NIM CRM GBW09202 Uric Acid |
| Quantity | Mass fraction |
| From (0) | 6 |
| To (0) | 200 |
| Unit (0) | µg/g |
| From (1) | 0.4 |
| To (1) | 2 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Peer Reviewed Publication | J. Chromatogr. B, 2007, 857, 287-295
|
| Additional Comments | There is no writing document available for the method that is totally compliant with the mandatory elements of ISO 15193. |
| Review Team Name | Metabolites and substrates |
| Analyte Category Name | Metabolites and substrates |
| Measurement principle/technique | Isotope dilution mass spectrometry |