JCTLM Database: higher-order reference materials, methods and services
Details : C7RMP2R - IFCC reference measurement procedure (37 °C) for Alkaline Phosphatase
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | C7RMP2R |
Analyte Name | alkaline phosphatase (ALP) |
IUPAC IFCC Number | EC3.1.3.1 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | lyophilized, fresh or frozen blood serum, or calibration solution or other biological materials |
Name | IFCC reference measurement procedure (37 °C) for Alkaline Phosphatase |
Organization that developed Validated Method | IFCC |
Measurement Techniques Used | kinetic spectrophotometry |
Quantity | Catalytic activity concentration |
From (0) | 0.05 |
To (0) | 11.4 |
Unit (0) | µkat/l |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Uncertainty Range | 3.0 % to [dependent from the matrix and the special properties of the analyte] % |
Peer Reviewed Publication | Clin. Chem. Lab. Med., 2011, 49(9), 1439-1446; IFCC Primary Reference Procedures for the Measurement of Catalytic Activity Concentrations of Enzymes at 37 °C
Part 9. Reference Procedure for the Measurement of Catalytic Concentration of Alkaline Phosphatase
|
Comments to be published via JCTLM DB | The catalytic activity concentration expressed in U/l ranges from 3.0 U/l to 685 U/l |
Analyte Category Name | Enzymes |
Measurement principle/technique | Spectrophotometry |