Accurate results
for patient care
Laboratory medicine and in vitro diagnostics

JCTLM Database: higher-order reference materials, methods and services

Details : C8RMP1 - University of Ghent reference measurement procedure for free thyroxine in serum

Type (acronym)rmp
Typereference measurement procedure
Unique Nomination NumberC8RMP1
Analyte Namefree thyroxine
IUPAC IFCC Numberfree thyroxine
Method capable of Traceability of Analyte to SI or Defined Proceduredefined procedure
Applicable Matriceslyophilized, fresh, or frozen serum
NameUniversity of Ghent reference measurement procedure for free thyroxine in serum
Organization that developed Validated MethodUniversity of Ghent
Measurement Techniques Usedequilibrium diaysis ID/LC/MS
Comments 1Equilibrium dialysis under well defined conditions followed by ID-LC-/tabdem MS measurement based on JCTLM NRMeth 11
Comments 2Calibrated by JCTLM listed T4 certified reference material IRMM 468, accuracy control with DGKL-sera certified for total T4 and self-prepared IQC pools
QuantityAmount-of-substance concentration
From (0)1.8
To (0)80
Unit (0)pmol/L
From (1)6.9
To (1)6.9
Unit (1)%
Level of Confidence (%)95%
Peer Reviewed PublicationClin. Chem. Lab. Med., 2011, 49(8), 1275 - 1281
Analyte Category NameNon-peptide hormones
Measurement principle/techniqueIsotope dilution mass spectrometry