Details : C8RMP4 - University of Ghent reference measurement procedure for 25-hydroxyvitamin D2
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | C8RMP4 |
Analyte Name | 25-hydroxyvitamin D2 |
IUPAC IFCC Number | CAS21343-40-8 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | lyophilized, fresh, or frozen serum |
Name | University of Ghent reference measurement procedure for 25-hydroxyvitamin D2 |
Organization that developed Validated Method | University of Ghent |
Measurement Techniques Used | ID/LC/MS |
Comments 1 | calibrated with NIST SRM 2972, accuracy control with NIST SRM 972 |
Quantity | Amount-of-substance concentration |
From (0) | 1.2 |
To (0) | 280 |
Unit (0) | nmol/L |
From (1) | 3.9 |
To (1) | 3.9 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | Clin. Chem., 2011, 57(3), 441-448
|
Analyte Category Name | Vitamins and micronutrients |
Measurement principle/technique | Isotope dilution mass spectrometry |