JCTLM Database: higher-order reference materials, methods and services
Details : C8RMP5 - CDC ID/GC/MS reference measurement procedure for cholesterol
Type (acronym) | rmp |
Type | reference measurement procedure |
Unique Nomination Number | C8RMP5 |
Analyte Name | total cholesterol |
IUPAC IFCC Number | CAS57-88-5 |
Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
Applicable Matrices | lyophilized, fresh, or frozen human serum |
Name | CDC ID/GC/MS reference measurement procedure for cholesterol |
Organization that developed Validated Method | Centers for Disease Control and Prevention |
Measurement Techniques Used | ID/GC/MS |
Comments 1 | trimethylsilyl derivative |
Comments 2 | Calibrate with SRM 911, high purity cholesterol |
Quantity | Amount-of-substance concentration |
From (0) | 1.1 |
Unit (0) | mmol/L |
Unit (1) | % |
Level of Confidence (%) | 95% |
Peer Reviewed Publication | Edwards SH, et al Clin. Chem., 2011, 57, 614-622
|
Analyte Category Name | Metabolites and substrates |
Non-CCQM Key Comparison Report | IFCC-RELA (External Quality assessment scheme for Reference Laboratories in Laboratory Medicine) |
Non-CCQM Key Comparison Report Hyperlink | http://www.dgkl-rfb.de:81/index.shtml |
Measurement principle/technique | Isotope dilution mass spectrometry |