JCTLM Database: higher-order reference materials, methods and services
Details : C8RMP6 - CDC definitive method for serum total testosterone
| Type (acronym) | rmp |
| Type | reference measurement procedure |
| Unique Nomination Number | C8RMP6 |
| Analyte Name | total testosterone |
| IUPAC IFCC Number | CAS58-22-0 |
| Method capable of Traceability of Analyte to SI or Defined Procedure | SI |
| Applicable Matrices | lyophilized, fresh, or frozen human serum |
| Name | CDC definitive method for serum total testosterone |
| Organization that developed Validated Method | CDC (Centers for Disease Control and Prevention) |
| Measurement Techniques Used | ID-LC/MS/MS |
| Comments 1 | Calibrated with ANMI M914 |
| Quantity | Amount-of-substance concentration |
| From (0) | 0.09 |
| To (0) | 34.7 |
| Unit (0) | nmol/L |
| From (1) | 2.9 |
| To (1) | 2.9 |
| Unit (1) | % |
| Level of Confidence (%) | 95% |
| Peer Reviewed Publication | Clin. Chem., 2013, 59(2), 372
|
| Analyte Category Name | Non-peptide hormones |
| Measurement principle/technique | Isotope dilution mass spectrometry |