Type (acronym) | rms |
---|---|
Type | reference measurement service |
Unique Nomination Number | C8RMS10 |
Reference Laboratory Status | Hospital |
Analyte Name | urea |
IUPAC IFCC Number | CAS57-13-6 |
Applicable Matrices | lyophilized, fresh or frozen human serum and calibration solution |
Traceable to SI | Yes |
Principle of Analytical Procedure | Spectrophotometry |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | CDC reference method for urea
|
JCTLM Database Identifier Number | NRMeth 36 |
Procedure Difference in Your Laboratory | No |
Quantity | Amount-of-substance concentration |
From (0) | 2.5 |
To (0) | 26.9 |
Unit (0) | mmol/L |
From (1) | 2.4 |
To (1) | 2 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Concentration Range | The measurement range spans from 2.5 mmol/L to 26.9 mmol/L without dilution of the sample. The range of concentration is adapted to the usual concentrations measured. |
Comments on Uncertainty Range | The expanded uncertainty values do not relate to the lower and higher limits of determination that are in accordance with the reference method. |
Reference Procedure Status Acceptable | Yes |
Laboratory Status | (b) Accredited (ISO 17025+15195) |
Survey Organisation | IFCC RELA |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Additional Information 1 | IFCC:RELA2010~2012 |
Participation Status in Intercomparison Surveys acceptable | Yes |
Comments of Review Teams | Recommended for Listing ! |
Organization Acronym | HT711 |
Organization Name | Beijing Aerospace General Hospital Reference Laboratory |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Metabolites and substrates |