JCTLM Database: higher-order reference materials, methods and services
Details : C8RMS11 - total protein in High purity material, Calibration solution, Blood serum, Blood plasma
Type (acronym) | rms |
Type | reference measurement service |
Unique Nomination Number | C8RMS11 |
Reference Laboratory Status | Hospital |
Analyte Name | total protein |
IUPAC IFCC Number | total protein |
Traceable to SI | Yes |
Principle of Analytical Procedure | Biuret reaction (Spectrophotometry) |
Reference Method Listed in JCTLM | Yes |
Name of JCTLM Listed Reference Method Procedure | Total serum protein,NRSCL RS5-A2 |
JCTLM Database Identifier Number | NRMeth 30 |
Procedure Difference in Your Laboratory | No |
Quantity | Mass concentration |
From (0) | 44.5 |
To (0) | 91.4 |
Unit (0) | g/L |
From (1) | 1.4 |
To (1) | 1.3 |
Unit (1) | % |
Level of Confidence (%) | 95% |
Comments on Concentration Range | The measurement range spans from 44.50 g/L to 91.40 g/L without dilution of the sample.
The range of concentration is adapted to the usual concentrations measured. |
Comments on Uncertainty Range | The expanded uncertainty values do not relate to the lower and higher limits of determination that are in accordance with the reference method. |
Reference Procedure Status Acceptable | RMP is Acceptable but has been delisted by JCTLM WG-1 because supporting documentation is not longer available |
Laboratory Status | Accredited with ISO 17025 and ISO 15195 (certificates supplied in supporting materials) |
Survey Organisation | IFCC,www.dgkl-rfb.de:81 |
Intercomparison Results Publicly Available | Yes |
Link to Comparison Results | RELA - IFCC External Quality assessment scheme for Reference Laboratories in Laboratory Medicine at AND http://www.dgkl-rfb.de:81/index.shtml |
Additional Information 1 | IFCC:RELA2009~2011 |
Participation Status in Intercomparison Surveys acceptable | Yes |
Comments of Review Teams | Although the RMP used for total proteins has been delisted from the JCTLM database, this is only due to lack of availability of the supporting documentation of the previously listed RMP |
Comments for JCTLM DB Publication | The primary calibrator used is a bovine albumin material. |
Organization Acronym | HT711 |
Organization Name | Beijing Aerospace General Hospital Reference Laboratory |
Organization Country ISO Code | CN |
Organization Country Name | China |
Analyte Category Name | Proteins |